CVE-2024-48966

CRITICAL
10.0
CVSS 3.1
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Metadata

CVE ID
CVE-2024-48966
State
PUBLISHED
Assigner
Baxter
Reserved
2024-10-10 19:24 UTC
Published
2024-11-14 21:38 UTC
Last updated
2024-11-15 15:37 UTC
Primary CWE
CWE-306
CWE-306 Missing Authentication for Critical Function
Vendor / Product
Baxter / Life2000 Ventilation System
Sources
cve.org  ·  NVD

Severity & Metrics

10.0 CRITICAL CVSS 3.1
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
SSVC — CISA Coordinator
Exploitation
none
Automatable
yes
Tech. Impact
total
Affected products (1)
VendorProductPlatformVersions
Baxter Life2000 Ventilation System 06.08.00.00 and prior
Weakness (CWE)
CWESourceDescription
CWE-306 cna CWE-306 Missing Authentication for Critical Function
CVSS scores (1)
ScoreSeverityVersionSourceVector
10.0 CRITICAL 3.1 cna CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H