Back to overview

CVE-2026-0611

CRITICAL Exploitation: PoC
9.8
CVSS 3.1
Description
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.

Metadata

CVE ID
CVE-2026-0611
State
PUBLISHED
Assigner
VulnCheck
Reserved
2026-01-05 16:55 UTC
Published
2026-06-02 15:39 UTC
Last updated
2026-06-02 19:13 UTC
Primary CWE
CWE-306
Missing Authentication for Critical Function
Vendor / Product
Spacelabs Healthcare / Sentinel
Sources
cve.org  ·  NVD

Severity & Metrics

9.8 CRITICAL CVSS 3.1
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
SSVC — CISA Coordinator
Exploitation
PoC
Automatable
yes
Tech. Impact
total
Affected products (1)
VendorProductPlatformVersions
Spacelabs Healthcare Sentinel 10.5.0 < 11.6.0
Weakness (CWE)
CWESourceDescription
CWE-306 cna Missing Authentication for Critical Function
CVSS scores (2)
ScoreSeverityVersionSourceVector
9.8 CRITICAL 3.1 cna CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
9.2 CRITICAL 4.0 cna CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Back to overview